Chip-based Digital PCR
In each digital PCR reaction, 1µl of 40 ng/µl genomic DNA was used and if it was a gblock oligo, the input was 1µl of 1aM of gblock oligo. In the total 20 µl of reaction, there were 8.5 µl of molecular grade water, 10 µl of 2X digital PCR Master Mix buffer from Thermo Fisher Scientific, 0.5 µl of 40X primer and probe, and 1µl of template. The procedures were performed according to the Quantstudio 3D digital PCR system guide from Thermo Fisher Scientific. In brief, the chip was placed on the 40°C heat plate for 5 minutes, and at the same time the reaction mix was incubated at room temperature for 10 minutes. 15 µl of reaction mix was loaded into the sample loading blade. The sample loading blade was held at a 70-80° angle and was dragged on the chip. The chip was processed for 20 seconds after loading the sample. Then, the chip was covered with the case lid, filled with immersion fluid, and sealed using the chip sealant with an UV stylus treatment for 30 seconds. The chip was loaded on the flat block GeneAMP 9700 PCR instrument with the cycle condition: 96°C for 10 minutes, 40 cycles of 98°C for 30 seconds, 60°C for 2 minutes, and 10°C for final resting temperature. After PCR, the chip was inserted into the Quantstudio 3D instrument for image analysis and the data was transferred to a USB drive. The data analysis was done through the Quantstudio 3D analysisSuite software.
Analytic sensitivity and linearity
Wild type and mutant template gblock oligos at 1 aM concentration were used to perform a mixing experiment and produce a series of KIT D816V mutation percentage dilutions from 100%, 10%, 1%, 0.1%, 0.01%, and 0.001% mutation. HMC-1.2 cell line DNA (40 ng/µl) which harbor heterozygous D816V mutation was diluted into D816V negative cell line HL-60 DNA with equal concentration. These included 100%, 80%, 50%, 30%, 10%, 1%, 0.1%, 0.01%, and 0.001% mutation dilution points. DPCR was performed on both series dilution from gblock oligos and cell line DNAs. The expected mutation percentage and dPCR quantitative mutation percentage were compared. The linearity and sensitivity were calculated based on these comparisons.
Patient validation, specificity, and reproducibility studies
Six SM patient samples with 40 ng of DNA input were used to perform digital PCR to detect KIT D816V. The percentage of the mutation from digital PCR was compared to Sanger sequencing results. The specificity test used six normal donor samples without any KIT D816V mutation to conduct the digital PCR assay. 3 samples from SM patients were chosen to run digital PCR for three times. The results were evaluated for reproducibility.
Results
Analytic sensitivity and linearity
Figure 1 shows a typical KIT D816V digital PCR result. The left side produced the chip image and right side produced each reaction well plot. Both gblock oligo and HMC 1.2 cell line digital PCR results are shown in Table 2. Gblock mutant oligo dilution with wild type gblock oligo correlated well between expected and experimental results. According to Figure 2, gblock dilution linearity was R2=0.9559. However, at the level of 0.01% dilution, the experimental result is 0.047%, which is about a difference. At the 0.001% level dilution, the experimental result is 0.018% ,which is almost one away from the expected percentage. Similar to the gblock oligo dilution, the HMC 1.2 cell line DNA dilution has linearity of R2=0.953, which is less than gblock dilution linearity R2 value. The last two dilutions also showed a big difference between expected percentage and experimental percentage—0.005% vs. 0.0281% and 0.0005% vs. 0.0104%.
Figure 1. A typical result display of chip based digital PCR detection KIT D816V
Left side is the image of the assay’s chip. Right side is the result plot based on the FAM and VIC fluorescent channel. The yellow background and dots show the empty wells without fluorescence. Blue dots indicate the FAM channel positive wells (D816V positive wells). Red dots show the VIC channel positive wells (D816V negative wells). Green dots show the mix of FAM and VIC channel wells. It counts as both negative and positive wells.
Table 2. Gblock Oligo and HMC-1.2 cell line DNA dilution dPCR Results

Figure 2. Linearity and Sensitivity of dilution of gblock oligo and HMC-1.2 cell line DNA
Gblock dilution (up) showed correlation of R2= 0.9559 which is close to the gblock oilgo dilution; the cell line DNA dilution (down) produced a correlation of R2= 0.953.
Patient validation
Six SM patient run results are shown in Figure 3. Patient No. 29 and 38 are positive in both dPCR and Sanger sequencing. No. 34 and 47 are negative in Sanger sequencing and positive in dPCR with 0.31% and 0.13%, respectively, due to lower sensitivity of Sanger sequencing. No. 28 and 36 are negative on both assays.
Figure 3. Six SM patients’ dPCR and Sanger sequencing comparison
Left number showed the SM patient number. The red arrow in middle of Sanger sequencing tracings showed the single base pair mutation (c. 2447 A toT). Right side showed digital PCR plots correspond the same SM patients. Right side percentages are the digital PCR calculation D816V mutation percentage.
Specificity and reproducibility
Six normal donor samples were used to perform dPCR and the results were 0.00%, 0.00%, 0.000775%, 0.00%, 0.00%, and 0.00%. One sample had a very low positive result. Three patients from validation were used to conduct reproducibility tests (Table 3). Patient No. 28 does not possess the D816V mutation and reproducibility test using three test runs produced all negative results. Patient No. 29 had the highest D816V mutation of 22.79%. A reproducibility run showed mutation percentages of 20.7%, 19.827%, and 21.74%, respectively. Statistics showed a standard deviation (SD) of 1.28 and a coefficient of variation (CV) of 6.03%. Patient No. 34 carried a 0.313% D816V mutation and the three reproducibility runs were tested. The experiments using Patient No. 34 had a 0.09 SD and a 28.22% CV.
Table 3. Three patients and three trial’s reproducibility
Discussion
Digital PCR characterized by high sensitivity and reproducibility has been rapidly adopted in clinical laboratory settings (Thermo 2016). There are two major digital PCR platforms: droplet digital PCR, such as Bio-Rad QX-200 system (ddPCR), and microchip based digital PCR (cdPCR), such as the Thermo Fisher Quantstudio 3D digital PCR system. Both platforms can achieve similar results in the detection of rare gene mutations (Huggett 2015). In 2018, the Hoermann group adopted ddPCR system to detect KIT D816V mutation with 0.01% sensitivity (Greiner 2018). To compare ddPCR with cdPCR, the chip based digital PCR system was used to detect KIT D816V mutation. Based on both the gblock oligo and the HMC 1.2 cell line dilution, the limit of detection is beyond 0.01%. Specifically, when the theoretical percentage is lower than 0.01%, the experimental percentage results are less accurate (Table 2). For example, at a 0.001% expected mutation value, the experimental result is 0.018%. The difference is that the percentage ratios at the 0.1% level are much smaller compared to the ratios at the 0.001% level. Discordance between the expected and experimental value within the 0.001% level may indicate a misrepresentation of mutation percentage, or the undetectable nature of the sample. In order to detect samples at the 0.001% level, an increased number of PCR wells, specifically on the order of 100,000 wells, are needed for a high-sensitive digital PCR. Since this chip only has 20,000 wells, the sensitivity detection at the 0.01% level indicates the presence of 1 positive well in every 10,000 wells. From these points, the cdPCR KIT D816V detection sensitivity or detection LOD is at 0.01%. This result is very similar to ddPCR limit of detection.
The specificity test of cdPCR was performed using normal donor DNA. Although normal donor DNAs do not carry the D816V mutation, there is one case that showed a very low detection for positive D816V mutation. Since it was well below the limit of detection at 0.01%, it was considered negative. The false detection of the mutation may be caused by a cdPCR reaction hypersensitivity or a digital PCR calculation error based upon Poisson distribution.
According to this study, digital PCR has excellent inter-run reproducibility. Both low and high positive samples have low standard deviation and CV% (< 30%). This reproducibility cannot be achieved using real-time PCR, which is around ½ log of precision.
Overall, the KIT D816V mutation detection using a 20,000 reaction well chip-based digital PCR had a very high sensitivity, linearity, and precision. Limit of detection of this assay is around 0.01%, which is similar to the droplet-based digital PCR system. This 0.01% LOD can be used as a tool for minimal residual disease (MRD) assays. To further extend LOD of cdPCR to 0.001% or lower, a higher density chip containing an increased number of wells to 100,000 or 1,000,000 wells is necessary.
References
- Cairoli, R. “Prognostic Impact of c-kit Mutations in Core Binding Factor Leukemias: An Italian Retrospective Study.” Blood 107.9 (2006): 3463-468.
- Edling CE and Hallberg B (2007). c-kit-a hematopoietic cell essential receptor tyrosine kinase. Int. J. Biochem. Cell Biol. 39 (11): 1995–1998.
- Greiner G, Gurbisz M, Ratzinger F, Witzeneder N, Simonitsch-Klupp I, Mitterbauer-Hohendanner G, Mayerhofer M, Müllauer L, Sperr WR, Valent P, Hoermann G. 2018. Digital PCR: A Sensitive and Precise Method for KIT D816V Quantification in Mastocytosis. Clin Chem. 64(3): 547-555.
- Huggett JF, Cowen S, Foy CA. Considerations for digital PCR as an accurate molecular diagnostic tool. Clin Chem 2015. 61:79 – 88.
- Jawhar M, Schwaab J, Naumann N, et al. (2017) Response and progression on midostaurin in advanced systemic mastocytosis: KIT D816V and other molecular markers. Blood. 130(2):137-145.
- Kim ES (2017) Midostaurin: First global approval. Drugs 2017 Jul;77(11): 1251-1259.
- Lawrence J.Jennings, David George, Juliann Czech, Min Yu, and Loren Joseph. (2014). Detection and quantification of BCR-ABL1 fusion transcripts by droplet digital PCR. Journal of Molecular Diagnostics. 16(2): 174-179.
- Pardanani A, Tefferi A: A critical reappraisal of treatment response criteria in systemic mastocytosis and a proposal for revisions. Eur J Haematol 2010, 84:371–378.
- Swerdlow SH, Campo E, Harris NL, Jaffe ES, Pileri SA, Stein H, Thiele J, Vardiman JW (Eds.): WHO classification of tumors of haematopoietic and lymphoid tissues. IARC: Lyon 2008, 54–63.
- Thermo Fisher Scientific 2016. Rare Mutation Analysis Using Digital PCR. N.p., n.d. Web. 27 Nov.
- Ustun C, DeRemer DL, and Akin C. (2011) Tyrosine kinase inhibitors in the treatment of systemic mastocytosis. Leukemia research 35:1143-1152.



